beautypg.com

2 specifications of las mkii receiver, 3 specifications of las mkii scintillometer system – Kipp&Zonen LAS MkII Scintillometer User Manual

Page 66

background image

66

Instruction Manual - LAS MkII Scintillometer

.

Note

Kipp & Zonen reserves the right to make changes to specifications without prior notice.

7.1 Specifications of LAS MkII transmitter

Parameter

Value

Operating temperature

-20 °C to +50 °C

Humidity

0 to 100 % RH, IP 65

Voltage

12 VDC nominal

(9.6 to 18 VDC)

Power

3 W maximum with heater off

51 W maximum with heater on, typical 23 W

Window heater

Self-regulating at approx. +55 °C

Surge protection

For transmitter power, but not for heater

Centre wavelength of optical filter for photodiode detector

850 nm

Fresnel lens diameter and focal length

152 mm / 152 mm

Scintillation signal bandwidth

0.2 to 400 Hz

Typical value signal 1

(

U

C

n

2

, log

C

n

2

signal)

0 to 2 V

Typical value signal 2

(6.5 - 7 kHz carrier)

0 to 1 V, 6.5 to 7 kHz, 0.5 duty cycle

Typical value signal 3

(

U

DEMOD

, signal strength

I)

0 to 2 V

C

n

2

turbulent range

10

-17

to 10

-12

m

-2/3

Noise level of electronics

< 10

-17

m

-2/3

Length

266 mm *

Width

229 mm *

Height

297 mm *

Weight

8.5 kg *

* Including pan and tilt assembly, but excluding sun shield or telescope

Parameter

Value for LAS MkII

Value with restrictors

Nominal aperture

150 mm

100 mm

Minimum path length

250 m

100 m

Maximum path length

(depends on atmospheric conditions)

4.5 km

1 km

Minimum height

1.5 m

1.5 m

Maximum height for

C

n

2

(or turbulence)

measurements

no operational restriction

no operational restriction

Maximum height for

H measurements Constant Flux Layer

Approx. 100 m

Approx. 100 m

7.2 Specifications of LAS MkII receiver

7.3 Specifications of LAS MkII scintillometer system