2 specifications of las mkii receiver, 3 specifications of las mkii scintillometer system – Kipp&Zonen LAS MkII Scintillometer User Manual
Page 66
66
Instruction Manual - LAS MkII Scintillometer
.
Note
Kipp & Zonen reserves the right to make changes to specifications without prior notice.
7.1 Specifications of LAS MkII transmitter
Parameter
Value
Operating temperature
-20 °C to +50 °C
Humidity
0 to 100 % RH, IP 65
Voltage
12 VDC nominal
(9.6 to 18 VDC)
Power
3 W maximum with heater off
51 W maximum with heater on, typical 23 W
Window heater
Self-regulating at approx. +55 °C
Surge protection
For transmitter power, but not for heater
Centre wavelength of optical filter for photodiode detector
850 nm
Fresnel lens diameter and focal length
152 mm / 152 mm
Scintillation signal bandwidth
0.2 to 400 Hz
Typical value signal 1
(
U
C
n
2
, log
C
n
2
signal)
0 to 2 V
Typical value signal 2
(6.5 - 7 kHz carrier)
0 to 1 V, 6.5 to 7 kHz, 0.5 duty cycle
Typical value signal 3
(
U
DEMOD
, signal strength
I)
0 to 2 V
C
n
2
turbulent range
10
-17
to 10
-12
m
-2/3
Noise level of electronics
< 10
-17
m
-2/3
Length
266 mm *
Width
229 mm *
Height
297 mm *
Weight
8.5 kg *
* Including pan and tilt assembly, but excluding sun shield or telescope
Parameter
Value for LAS MkII
Value with restrictors
Nominal aperture
150 mm
100 mm
Minimum path length
250 m
100 m
Maximum path length
(depends on atmospheric conditions)
4.5 km
1 km
Minimum height
1.5 m
1.5 m
Maximum height for
C
n
2
(or turbulence)
measurements
no operational restriction
no operational restriction
Maximum height for
H measurements Constant Flux Layer
Approx. 100 m
Approx. 100 m
7.2 Specifications of LAS MkII receiver