1.1 Product overview
The LAS MkII Large Aperture Scintillometer is an optical instrument with a 150 mm diameter beam that is designed for measuring
the path-averaged structure parameter of the refractive index of air (C
n
2
) over horizontal path lengths from 250 m to 4.5 km
1
. When
the supplied 100 mm diameter aperture restrictors are fitted, the path length can be from 100 m to 1 km
1
.
LAS MkII uses a transmitter and receiver horizontally separated by several kilometres to measure intensity fluctuations in the
air known as scintillations. This is the same effect, but of much smaller amplitude, as the ‘shimmering’ of air over very hot or
cold surfaces that causes a mirage.
The scintillations seen by the instrument can be expressed as the structure parameter of the refractive index of air (C
n
2
). The
light source of the LAS MkII transmitter operates at a near-infrared wavelength of 850 nm. At this wavelength the observed
scintillations are primarily caused by turbulent temperature fluctuations.
Therefore, C
n
2
measurements obtained with the LAS MkII can be combined with temporally and spatially coherent meteorological
observations of air temperature, wind speed and air pressure to derive the free convection sensible heat flux (H
free
). An accessory
meteorological sensor kit is available for this purpose, which connects to the LAS MKII receiver.
The LAS MkII is self-contained. The system can be locally configured at the receiver with a display and menu keys and has
internal digital signal and data processing and data storage. When the data is exported to the included EVATION software
package running on a computer the surface sensible heat flux (H) can be calculated.
Compared to traditional point measurement systems, the LAS MkII operates at spatial scales comparable to the grid box size of
numerical models and the pixel size of satellite images used in meteorology, hydrology and water management studies. The LAS MkII
has important applications in energy balance and water balance studies, because the surface flux of sensible heat is linked to latent
heat flux (L
v
E
) and evapotranspiration (ET). For these measurements a complete LAS MkII ET System is available.
This manual provides information related to the installation, maintenance, calibration, product specifications and applications
of the scintillometer.
If any questions should remain, please contact your local Kipp & Zonen representative or e-mail the Kipp & Zonen customer and
product support department at:
Please go to
www.kippzonen.com for information about other Kipp & Zonen products, or to check for any updates to this manual
or software.
1
The maximum usable path length depends upon the atmospheric conditions, the path lengths given are for ‘clear’ conditions (visibility 10-20 km). In general, it is
best to use the scintillometer at full aperture at path lengths down to 250 m, and the aperture restrictors for shorter paths. However, for field-scale measurements
at a range of distances up to 1 km it may be convenient to leave the restrictors fitted.
3
Instruction Manual - LAS MkII Scintillometer
.
Dear customer, thank you for purchasing a Kipp & Zonen instrument. It is essential that you read this manual completely for a
full understanding of the proper and safe installation, use, maintenance and operation of your new LAS MkII Scintillometer.
We understand that no instruction manual is perfect, so should you have any comments regarding this manual we will be
pleased to receive them at:
Kipp & Zonen B.V.
Delftechpark 36, 2628 XH Delft, - or
P.O. Box 507, 2600 AM Delft,
The Netherlands
T: +31 (0) 15 2755 210
F: +31 (0) 15 2620 351
[email protected]
www.kippzonen.com
Warranty and liability
Kipp & Zonen guarantees that the product delivered has been thoroughly tested to ensure that it meets its published specifications.
The warranty included in the conditions of delivery is valid only if the product has been installed and used according to the
instructions supplied by Kipp & Zonen.
Kipp & Zonen shall in no event be liable for incidental or consequential damages, including without limitation, lost profits, loss
of income, loss of business opportunities, loss of use and other related exposures, however incurred, rising from the faulty and
incorrect use of the product.
Modifications made by the user may affect the instrument performance, void the warranty, or affect the validity of the CE
declaration or other approvals and compliances to applicable International Standards.
Copyright © 2012 Kipp & Zonen B.V.
All rights are reserved. No part of this publication may be reproduced, stored in a retrieval system or transmitted, in any form
or by any means, without authorisation by Kipp & Zonen.
Kipp & Zonen reserves the right to make changes to this manual, brochures, specifications and other product documentation
without prior notice.
Manual document number: V1207
Publication date:
1
st
July 2012
Important User Information
1.2 Key parts of the LAS MkII Scintillometer
1.2.1 Transmitter and receiver
The drawing shows the key common parts of the LAS MkII transmitter and receiver:
Sun shield fitted to the mounting for the alignment telescope
Tilt (vertical) adjustment screws
Pan (horizontal) adjustment screws
Transmitter window, with heater and Fresnel lens behind
Drying cartridges
Baseplate
1.2.2 Transmitter rear panel
The drawing shows the key parts of the LAS MkII transmitter rear panel:
Power indicator (red)
Transmitter power adjustment knob (remove screw-on cover)
Signal output connector (4-pin)
Power input connector (12 VDC)
.
1.2.3 Receiver rear panel
The drawing shows the key parts of the LAS MkII receiver rear panel:
Display
Power and status indicator (green)
Menu navigation keys
Meteorological sensor kit connector (8-pin)
Analogue signal connector (4-pin)
Power input connector (12 VDC)
Digital interface connector (8-pin)