All the instruments are factory calibrated for temperature.
Hanna’s temperature probes are interchangeable and no temperature
calibration is needed when they are replaced.
If the temperature measurements are inaccurate, temperature recalibration
should be performed.
For an accurate recalibration, contact your dealer or the nearest Hanna
Customer Service Center, or follow the instructions below.
• Prepare a vessel containing ice and water and another one containing

hot water (around 50 ºC). Place insulation material around the
vessels to minimize temperature changes.

• Use a calibrated thermometer with a resolution of 0.1 ºC as a

reference thermometer. Connect the EC probe to the appropiate
socket.

• With the instrument off, press and hold down

CFM&SETUP keys

and then power on the instrument. The “

CAL” tag will appear and

the secondary LCD will show “

0.0 ºC“.

• Submerse the EC probe in the vessel with ice and water as near as

possible to the calibrated thermometer. Allow a few seconds for the
probe to stabilize.

• Use the

ARROW keys to set the reading on the secondary LCD to that

of ice and water, measured by

the calibrated thermometer. When

the reading is stable and close to the selected calibration point, the

CFM” tag will blink.

• Press

CFM to confirm. The secondary LCD will show

50.0

º

C

.

• Submerse the EC probe in the second vessel as near as possible to

the calibrated thermometer. Allow a few seconds for the probe to
stabilize.

TEMPERATURE CALIBRATION

(for technical personnel only)