For NaCl Range
• The conductivity and temperature reading value as described in TDS
Range.
• The time and date as described in EC Range.
• The
salinity factor on the primary LCD and the log number on the
secondary LCD, with “
RCL” and “NaCl” tags displayed.
• The reference temperature, the temperature compensation mode
and the cell constant message as described above.
Note: When an information that does not display the record number
is selected, pressing the
SETUP key will display the record
number on the secondary LCD line.
To delete logged records press the
CLR key,while viewing the logged
data. The “
dEL” message is displayed on the primary LCD and the
selected record on the secondary LCD, along with ”
RCL” tag.
• The
ARROW keys can be used to change the record number.
• Press
SETUP to select delete all records feature. The display will show
“
dEL” in the primary LCD and “ALL” in the secondary LCD.
• Press
CFM to confirm delete. While deleting the “ ” tag will blink.
• Press
CAL or RANGE or CLR to escape and return to the RCL screen.
• If “
dEL ALL” option was selected, all the log on demand records are
deleted. While deleting the “ ” tag is displayed blinking.
• Press
ALT&RCL to return to measurement mode at any time.
• If one or more records were deleted the “ ” tag blinks until the log
memory space is reorganized.