CELL CONSTANT
Press
CAL when the cell constant is displayed. The cell constant value
and “
CFM” tag will start blinking.
Press the
ARROW keys to change the cell constant (0.500 to 1.700).
Press
CFM to save the modified cell constant.
Press
CAL to escape without saving.
TDS FACTOR
Press
CAL when “TDS fact.” is displayed. The TDS factor and “CFM”
tag will start blinking.
Press the
ARROW keys to change the TDS factor (0.40 to 0.80).
Press
CFM to save the modified TDS factor.
Press
CAL to escape without saving.
TEMPERATURE COMPENSATION COEFFICIENT
Press
CAL when the temperature compensation coefficient is displayed.
The temperature compensation coefficient and “
CFM” tag will start
blinking.
Press the
ARROW keys to change the temperature compensation
coefficient. (0.00 to 6.00 %/ºC).
Press
CFM to save the modified temperature compensation coefficient.
Press
CAL to escape without saving.
REFERENCE TEMPERATURE
Press
CAL when the reference temperature is displayed. The reference
temperature value and “
CFM” tag will start blinking.