• The temperature coefficient.

• The instrument ID.

NaCl CALIBRATION DATA

NaCl calibration data is stored automatically after a successful calibration.
To view the NaCl calibration data, press

GLP when the instrument is in

NaCl measurement mode.
The instrument will display the time (hh:mm) of the last calibration.

Use the

ARROW key to scroll through the calibration data.

• The date (mm.dd.yyyy).

• The salinity coefficient.

• The cell constant.

• The instrument ID.