Hanna Instruments HI 2300 User Manual

Page 11

background image

11

Notes: • If the reading is too far from the expected value, the

WRONG” and “ ” tags will blink. Calibration can not be

confirmed.

In this case check if the calibration solution has been used

correctly or clean the probe by following the Probe Maintenance
(see page 31).

If the meter is in

Atc or Mtc mode and the temperature is

out of the 0.0 to 60.0 ºC interval, “

WRONG”, “ºC” tags

and the temperature will blink.

For best results choose an EC standard value close to the

sample to be measured.

In order to minimize any EMC interference, use plastic or

glass beakers.

It is possible to set the cell constant value directly, without

following the calibration procedure. To set the cell constant,
enter SETUP mode and select “

CELL” (see SETUP for details,

page 21).