beautypg.com

Mechanical dimensions, Specifications – Hanna Instruments HI 504 User Manual

Page 6

background image

11

10

Alarm Relay

Electromechanical Relay SPDT contact output,
5A - 250 VAC, 5A - 30 VDC (resistive load)
Fuse protected: 5A, 250V Quick Blow Fuse

Analog Output

• 2 independent outputs

• 0 - 22 mA (configuring as 0-20 mA or 4-20 mA)

Analog Output Resolution

0.1‰ f.s.

Analog Output Accuracy

± 2‰ f.s.

Data logging

6000 pH/°C or ORP samples

Environment

0 to 50 ºC; max 85% RH non-condensing

Casing

IP20 (housing); IP54 (front panel)

Enclosure

single case ½ DIN

Weight

approximately 1.6 kg (3.5 lb.)

MECHANICAL DIMENSIONS

FRONT VIEW

SIDE VIEW

SPECIFICATIONS

Range

-2.00 to 16.00 pH
-2000 to 2000 mV
-30 to 130.0 ºC

Resolution

0.01 pH
1 mV
0.1 ºC above -10 ºC; 1 ºC below

Accuracy

±0.02 pH

(@20°C/68°F)

±2 mV
±0.5 ºC (-9.9 to 130.0 ºC); ±1 ºC (-30 to -10 ºC)

Typical EMC Deviation

±0.2 pH
±10 mV
±0.5 ºC

Digital Input for

RS485

the pH/ORP/°C Transmitter

Other Digital Insulated Inputs • 2 digital insulated inputs: 1 for hold and 1 for

the advanced cleaning

• ON state: 5 to 24 VDC

Digital Insulated Output

A digital insulated contact closed upon hold mode

Temperature compensation Automatic or manual from -30 to 130 ºC

Temperature probe

With 3-wire or 2-wire Pt100/Pt1000 sensor
(with automatic recognition and damage test)

Installation Category

II

Power Supply

24 Vdc/ac

(depending on model)

230 Vac, 115 Vac or 100 Vac ±10%; 50/60 Hz

Power Consumption

10 VA

Over Current Protection

400 mA 250V Quick Blow Fuse

Max. Oscillation Frequency 8 MHz

Relays 1, 2, 3, 4

Electromechanical relay SPDT contact outputs,
5A-250 VAC, 5A - 30 VDC (resistive load)
Fuse protected: 5A, 250V Quick Blow Fuse