ELECTRODE MAINTENANCE

PREPARATION

Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT. This is normal with
electrodes and they will disappear when rinsed
with water.
During transport tiny air bubbles may form in-
side the glass bulb, and the electrode cannot
function properly under these conditions. Remove
the bubbles by "shaking down" the electrode
as you would do with a glass thermometer.

Reference
Filling Hole

Reference
Filling Hole

Fill Hole
Screw

Sensitive

Wire

Reference

Wire

Reference

Junction

Glass

Bulb

Glass
Bulb

Reference
Junction

Reference
Junction

Reference

Junction

Platinum or
Gold tip

Platinum or

Gold tip

Reference
Wire

Reference
Wire

Sensitive
Wire

Glass Body

pH Electrode

Plastic Body

pH Electrode

Plastic Body

ORP Electrode

Glass Body

ORP Electrode

Reference

Wire

Generally, if the ORP (mV) reading corre-
sponding to the pH solution value is higher
than the value in the table below, an oxidizing
pre-treatment is necessary; otherwise a re-
ducing pre-treatment is necessary:

pH mV pH mV pH mV pH mV pH mV

0 990 1 920 2 860 3 800 4 740

5 680 6 640 7 580 8 520 9 460

10

400 11

340 12

280 13

220 14 160

Reducing pre-treatment: immerse the elec-
trode for some minutes in HI 7091 solution.

Oxidizing pre-treatment: immerse the elec-
trode for some minutes in HI 7092 solution.

If no pre-treatment is performed, the elec-
trode will have long response times.

If working with refillable electrodes, always
check the internal electrolyte level and refillwith
HI 7071 solution, if necessary (the level must
be at least 2.5 cm below the filling hole).

If measurements are taken in solutions con-
taining sulfides or proteins, the cleaning of
the electrode junction must be performed (see
"Cleaning Procedure" section for details).

To check the correct functioning of the ORP
electrode, immerse it into HI 7020 test
solution and verify that the reading is within
200 and 275 mV.

After the test, rinse the electrode thoroughly
with water and proceed with the oxidizing or
reducing pre-treatment before taking any
measurement.

When not in use, the electrode tip should be
kept moist and far from any type of mechani-
cal stress which might cause damage. For
this reason, it is reccommended to store the
electrode with a few drops of HI 70300 storage
solution in the supplied protective cap.