OPERATIONAL & DIAGNOSTIC CODES

Code

Function

1 0 0

Increase recorder output lower limit in manual
low-scale adjustment

101

Decrease recorder output lower limit in
manual low-scale adjustment

1 0 2

Increase recorder output upper limit in
manual full-scale adjustment

1 0 3

Decrease recorder output upper limit in
manual full-scale adjustment

1 0 4

Set recorder 0 V automatically with dip
switch setting at 0-1 V, 0-100 mV or
0-10 mV

1 0 5

Set recorder full-scale automatically with
dip switch setting at 0-1 V, 0-100 mV or
0-10 mV

1 0 6

Set recorder 4 mA automatically with dip
switch setting at calibrate current

1 0 7

Set recorder 20 mA automatically with dip
switch setting at calibrate current

1 1 0

Show recorder output low-scale setting on
DVM or ammeter

111

Show recorder output full-scale setting on
DVM or ammeter

1 1 2

Show recorder output mid-range value on
DVM or ammeter

4 0 0

Set delta of proportional dosing

4 0 1

Show current delta of proportional dosing

5 0 0

LED’s test; momentarily turn on all display
LED segments, mg/L LED, dosing LED,
alarm LED and system error LED

5 0 1

Alarm relay test; momentarily energize dos-
ing relay, alarm relay and system error
relay

5 0 2

Prime peristaltic pump for 3 minutes or
until user presses the DIAG key