Hanna Instruments PCA 300 Series User Manual

Page 48

background image

48

DIAG

RELAY TEST

In the relay test mode all three alarm relays close for about
one second.

1. Press DIAG . The display will show the last diagnostic

mode and the right-most numerical digit will blink.

2. Enter code 501 using the and keys. After the code is

selected display will show “d501”.

3. Press DIAG. The display will blink “ON” and “OFF”, then

the display will automatically return to the last concentra-
tion reading.

CALIBRATION PROCEDURE

1. Withdraw a sample of the measured liquid direct from the

drain port of the measuring cell (#11) by opening its valve
(#12).

Note:

withdraw the sample just before the electrovalve
stops the liquid flow to the measuring cell.

2. With a calibrated meter take a measure of

the sample. This is the calibration value.