voltage limits can be calibrated automatically as follows:

1.A To automatically calibrate "ZERO" scale

1.A.1 Press DIAG. The display will show the last diagnostic

mode and the right-most numerical digit will blink.

Note

If the analyzer has been reset or turned on, “d000” will ap-
pear when pressing DIAG.

1.A.2 Enter code 104 using the and keys.

1.A.3 Press DIAG; “rc 0” will be displayed and the analyzer

will execute this mode.

1.B To automatically calibrate "FULL" scale

1.B.1 Press DIAG. The display will show the last diagnostic

mode and the right-most numerical digit will blink.

1.B.2 Enter code 105 using the and keys.

DIAG

DIAG

DIAG