Hanna Instruments PCA 300 Series User Manual
Page 21

21
Set the individual micro-switch
for the desired output in the ON
position (e.g. 0-1V).
See table below.
Switch
Function
6
Not used
5
Calibrate Current
4
4 - 20 mA
3
0 -1 V
2
0 -100 mV
1
0 -10 mV
Dosing Pump
Proportional dosing can be performed
connecting an external pump to the
DOSING RELAY block terminal.
For the correct connection, see the drawing
aside.