SPECIFICATIONS

PCA 300

PCA 301

Range

0.00 to 5.00 mg/L

0.00 to 5.00 mg/L

Free Cl

2

Total Residual Cl

2

Resolution

0.01 mg/L

Accuracy

±8% of reading or ±0.05 mg/L
whichever is greater

Typical EMC

±0.05 mg/L

Deviation

Minimum

0.05 mg/L

detectable level

Repeatability

±0.05 mg/L

Response time

Related to the sampling time selected.
Typical for a full scale step change and
5 minutes between two consecutive
samples: one sampling cycle for 90%
response and two sampling cycles for
100% response

Sampling rate

Adjustable from 3 to 102 minutes

Sample inlet

0.07 bar (1psig) min., 4 bar (57.2 psig)

pressure

max. An internal regulator reduces
pressure from 4 bar (57.2 psig) to
1 bar (14.3 psig)

Sample flow

Flow rate of 300 mL/min is recom-

range

mended. Minimum and maximum
allowed are 100 mL/min and
500 mL/min, respectively

Sample

5 to 40°C (41 to 104°F)

temperature range

Interferences

Oxidizing agents such as: Iodine,
Bromine, Ozone, Chlorine Dioxide,
Permanganate, Hexavalent Chromium.
Hardness must not exceed 1000 mg/L
as CaCO

3

. Alkalinity must not exceed