beautypg.com

Hanna Instruments HI 84502 User Manual

Page 41

background image

41

STORAGE PROCEDURE

To minimize clogging and assure a quick response time, the glass bulb and the junction of the
electrode should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of

HI 70300 Storage Solution or, in

its absence, with

HI 7082 Fill Solution. Follow the Preparation Procedure before taking

measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.

PERIODIC MAINTENANCE

Inspect the electrode and the cable. The cable used for connection to the instrument must be
intact and there must be no points of broken insulation on the cable or cracks on the electrode
stem or bulb. Connectors must be perfectly clean and dry. If any scratches or cracks are present,
replace the electrode. Rinse off any salt deposits with water.

PROBE MAINTENANCE

Refill the reference chamber with fresh electrolyte (

HI 7082). Allow the electrode to stand

upright for 1 hour. Follow the Storage Procedure above.

pH CLEANING PROCEDURE

Wine deposits

Soak in Hanna

HI 70635 cleaning solution for 15 minutes

Wine stains

Soak in Hanna

HI 70636 cleaning solution for 15 minutes

IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly
with distilled water, refill the reference chamber with fresh electrolyte and soak the electrode in
HI 70300 Storage Solution for at least 1 hour before taking measurements.