Titration procedure – Hanna Instruments HI 84502 User Manual

Page 28

background image

28

TITRATION PROCEDURE

• For best accuracy, before taking any measurement, ensure that the pump is calibrated on the

selected range following the “Pump Calibration Procedure” (see page 26).

Note: Verify that the instrument has been calibrated (pH and pump) before performing any

titrations.

• Refer to “Setup Menu” (see page 11) to set up the instrument for your measurement.
• Select the corresponding range according to the table below:

Sample preparation: Use a clean tip and 2000 uL automatic pipette to precisely add the
appropriate amount of wine sample to a clean beaker as indicated below:

Note: Failure to use a clean pipette will result in erroneous readings.

• Fill the beaker up to the 50 mL mark with distilled or deionized

water.

• Press

Titrator.

Note: DO NOT PLACE THE TIP INTO THE SAMPLE

BEAKER. PLACE THE TIP OVER A WASTE BEAKER.
A SMALL AMOUNT OF TITRANT IS DISPENSED
WHEN THE PUMP RESETS.

• Press

Start to begin a titration.

• Place the stir bar in the beaker and put the beaker in the

minititrator top.

• Place the probe holder on the top of the beaker and secure it

by turning clockwise.

• Rinse the pH electrode with deionized water and immerse

into the sample until the PTFE reference junction is completely
submerged. Be sure that the tip of the electrode is not hitting
the stir bar.