TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be followed carefully to ensure measurements are conducted
with the highest possible accuracy and precision.
• IT IS CRITICAL THAT THE TIP BE IMMERSED IN THE SOLUTION BEING TITRATED

(APPROXIMATELY 0.25 CM).

• Use a clean, volumetric pipette to measure and transfer the necessary volume of wine sample

into the titration beaker.

• Calibrate the pump prior to each series of titrations.
• Calibrate the pump if the meter is left idle for several hours.
• Analyze the wine sample immediately after it is obtained.
• Clean the electrode with

HI 700635 or HI 700636 cleaning solutions specially designed for

the wine industry.

• Press

Restart to begin a new titration or ESC to return to the titration menu.

• If the concentration exceeds the range limits

(> 5.0 g/L for

Low Range, > 25.0 g/L for

High Range) the exceeded range limit will be
displayed blinking. Another titration can be
started by pressing

Restart.

“Wrong input” error message appears when
the input reading (pH, temperature) exceeds the
specified limits. The pH or temperature value
and the concentration will blink indicating an
error.

• This screen appears when the stirrer is not

working properly. Check the stir bar and beaker
content. Press

Restart to try again.

• This error message appears when the pump is not

working properly. Check the tubing, valve and
syringe. Press

Restart to try again.