pH MEASUREMENT

The

HI 84502 can be used as a pH meter for direct measurements.

Verify that the instrument has been calibrated before taking pH measurements. Set the
instrument to

pH meter. From titrator mode press Meter until pH units are displayed.

If an electrode calibration hasn’t been performed, or the number of days exceeds the calibration
time out value set, the message

“CAL DUE” will blink on the left side of the display (see

Calibration timeout option in Setup for details).
If

CAL DUE is displayed perform an electrode calibration.

Press

MENU to access the instrument’s menu.

Press

HELP to view the contextual help, every time you need additional information.

Press

STIR to start/stop the stirrer.

Press

Titrator to enter titration mode.

Press

CAL to access the calibration menu.

Press

LOG to save the current reading. A message indicating the free log space will be displayed

for a few seconds.

Press

Pump to view the pump’s last calibration time, date and slope.

If a calibration hasn’t been performed, the message

“Not Calibrated” will be displayed.