beautypg.com

Electrode conditioning and maintenance – Hanna Instruments HI 84502 User Manual

Page 40

background image

40

ELECTRODE CONDITIONING AND MAINTENANCE

PREPARATION PROCEDURE

Remove the protective cap of the pH electrode (

HI 1048B).

DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. They will
disappear when rinsed with distilled/deionized water.
During transport, tiny bubbles of air may form inside the glass bulb, affecting proper functioning
of the electrode. These bubbles can be removed by “shaking down” the electrode as you would
do with a glass thermometer.
If the bulb and/or junction is dry, soak the electrode in

HI 70300 Storage Solution for at least

one hour.
If the fill solution (electrolyte) is more than 2½ cm (1”) below the fill hole, add

HI 7082 3.5M

KCl Electrolyte Solution.
For faster response, unscrew the fill hole screw during measurements.