Setup menu – Hanna Instruments HI 84502 User Manual

Page 11

background image

11

SETUP MENU

The titrator’s setup menu may be accessed from the main screen (meter or titrator) by pressing
the

MENU key, then Setup.

A list of setup parameters will be displayed with currently configured setting.
While in the setup menu it is possible to modify the instrument’s operation parameters. The
ARROW keys permit the user to scroll the setup parameters.
Press

HELP to view the contextual help.

Press

ESC to return to the main screen.

Range Setup

Use

Low measurement range for 0.1 - 5.0 g/L.

Use

High measurement range for 4.0 - 25.0 g/L.

To ensure a high accuracy, it is recommended to
recalibrate the pump after the valve, titrant or
electrode has been changed.

Use the

ARROW keys to select the value.

Press

Select to confirm or ESC to return to the setup

menu without saving the changes.

Calibration Timeout

Disabled or 1 to 7 days.
This option is used to set the number of days before
the pH calibration expired warning message is
displayed.
Press

Modify to access the calibration timeout screen.

pH Resolution

0.1 or 0.01.
Press the displayed virtual option key to change the
option.

7.00 or 8.20.
Press the displayed virtual option key to change the
option.

pH end point