Hanna Instruments HI 84502 User Manual

Page 20

background image

20

Point 1 calibration
• The default buffer selected will be 7.01 if end point was set to 7.00 pH or else 8.20 pH will

be selected. If necessary, press the

ARROW keys in order to select a different buffer value.

• The

(unstable measurement) symbol will be shown on the display until the reading

becomes stable.

• When the reading is stable and close to the selected buffer, the

(unstable measurement)

symbol will disappear and the

Confirm key will become active.

• Press

Confirm to confirm the calibration or ESC to exit calibration.

• After the first calibration point has been confirmed, press

ESC to exit without performing the

second calibration point.

Point 2 calibration
• The calibrated value will be shown on the display and the second expected buffer value will

be displayed.

• Remove the electrode holder with electrodes from the top of the beaker.
• Place the second beaker with calibration buffer in the beaker holder. Rinse the electrodes in

a beaker containing the second buffer rinsing solution.

• Place the electrode holder (with electrodes) on the top of the beaker and secure it by turning

clockwise and press

STIR.

• If necessary, press the

ARROW keys in order to select a different buffer value.