Hanna Instruments HI 84502 User Manual

Page 27

background image

27

Note: • If temperature probe is not connected, Manual Temperature Compensation is used

and

MTC appears on the right side of the screen. If Automatic Temperature

Compensation is in use the ATC appears on the right side of the screen.

• If an erroneous situation is encountered

during the calibration, an error message
is displayed and the calibration can be
restarted by pressing

Restart. Prepare a

new standard, rinse electrode, temperature
probe and dosing tip and try again.

• If the calibration doesn’t complete and

the max titrant volume of titrant is reached,
an error message will be displayed. The
calibration can be restarted by pressing
Restart. Prepare a new standard, rinse
electrode, temperature probe and dosing
tip and try again.

• This error message appears when the pH

reading exceeds the acceptable input limits
(-2.00 < pH < 16.00).

• This screen appears when the stirrer is

not working properly. Check the stir bar
and beaker content. Press

Restart to try

again.

• This error message appears when the pump

is not working properly. Check the tubing,
valve and syringe. Press

Restart to try

again.

• Insert the dosing tip into the titrant tube sleeve. IT IS CRITICAL

THAT THE TIP BE IMMERSED APPROXIMATELY 0.25 CM (0.1")
INTO THE SOLUTION BEING TITRATED.

• Press

Continue to begin the calibration and Stop to abort it.

• At the end of the calibration, ”

Calibration Completed“

appears on display. To repeat the calibration press

Restart or

ESC to return to the main screen.