SPECIFICATIONS

HI9218 HI9219

Ranges

pH

0.00 to 14.00

°C

-10.0 to 125.0

Resolution pH

0.01

°C

0.1

Accuracy

pH

±0.01

(

@

20°C/68°F) °C

±0.5

Typical EMC pH

±0.04

Deviation

°C

±1.5

Calibration

2 point with automatic recognition of

3 buffers (pH 4.01, 7.01, 10.01)

Temperature

-10° to 125°C (14 to 257°F) Completely

Compensation

automatic with HI 7669/2W temperature

probe; manual by means of keyboard

Comp. Interface

RS232C

Electrode

HI9218: FC100B plastic body combination

pH electrode (included)

HI9219:HI1131B Glass body combination

pH electrode (included)

Temperature

HI7669/2W Temperature probe (included)

Probe

Input Impedance

10

12

Ohm

Display

2 large 4-digit LCD for pH and temperature

with symbols 1 small LCD for date/time and

stability indicator

Printer

1 x 16 line dot matrix printer,
paper width: 44 mm,

roll diameter: 35 mm

Power Supply

Power socket for 12VDC

Environment

0 to 50°C (32 to 122°F); RH 95%

Dimensions

230x170x70mm (9.1x6.7x2.7")

Weight

1.3 Kg (2.9 lb.);

kit w/holder 3.2 Kg (7 lb.)

DATE/TIME key to display date and time.
CAL key

to enter calibration mode.

CFM key

to confirm memorized buffers 7.01 and 4.01/10.01.

UP/DOWN key

to set manual temperature compensation and date/time.

PAPER key

manual paper feed control.

PRINT key

prints pH value together with temperature, date
and time.

READY key

stability indicator to ensure correct pH reading or
to enable calibration.

This manual is related to the following products: