•• After about 30 seconds, the small display will

show "YES": press the CFM key to confirm the
calibration.

The LCD will show "Err5" to indicate that the electrode is now
ready for slope calibration and another step of instruction proce-
dures will be printed.

••

•• Rinse and immerse the pH electrode and tem-

perature probe in pH 4.01 or pH10.01 solution.
The LCD shows the corresponding buffer value at the working
temperature. If "Err5" does not disappear this means that the
buffer is not recognized as pH4 or pH10.

••

•• After about 30 seconds, the small LCD confirms the reading by

displaying "YES".

• Press the CFM key and the printer will confirm

that the pH calibration is completed.

Now dip the

electrode in 2nd

solution, wait

for 'YES' to

appear and then

press CFM

ºC

pH

CFM

°C

4 cm

(1½

")

HI 7004

ºC

pH

CFM

Calibration is

now completed.