beautypg.com

Hanna Instruments HI 9017 User Manual

Page 30

background image

30

e.g. 6.95 at 5°C:

• press 6
• press decimal point
• press 9
• press 5
• press ENTER A

Value 6.95 appears on the left display with pH symbol blinking. If
there is a difference of 1 pH between the two displays, there will be a
long beep.
Otherwise,

• press the CON key.

The left display value is transferred to the right display and the left
display shows "P1-2".

Rinse the electrode and dip it with the temperature
probe into the second solution HI7009 or HI8009
(pH 9.18), the value of which for example is 9.39
at 5°C (read from the standard charts). Wait for 30
seconds.

• press 9
• press decimal point
• press 3
• press 9
• press ENTER B

The value "9.39" appears on the left display with a blinking pH
symbol. If there is a difference of 1 pH between the two displays, there
is a long beep.
Otherwise,

• press the CON key.

ENTER

A

6

.

9

5

pH

pH

CON

pH

4 cm

(1½

")

HI 7009

ENTER

B

9

.

3

9

pH

pH

CON

This manual is related to the following products: