For the rinsing process, it is recommended to use a liberal amount of
the next solution to be measured.
Using refillable electrodes
Unscrew the small refill cap to open the refill hole during measure-
ments. After measurements replace the refill hole cap.
This will increase the pressure and give a faster response.

FOR HI 9218, HI 9219, HI 9318 ONLY

Press the PRINT key to get a printout at any time.
All the printouts will be complete with date, time,
sample No. (for HI9318 only), temperature and pH
(or mV with HI9318 only) value.

HI9218/HI9219

HI9318

Sample Printouts

MANUAL TEMPERATURE COMPENSATION

When the temperature probe is not connected, take the temperature of
the solution with a ChecktempC or a glass thermometer.

FOR HI 9017 ONLY

To manually set a temperature value (e.g. 37.2°C)
proceed as follows:
• press 3
• press 7
• press the key displaying decimal point "."
• press 2
• press TEMP

DATE

: 01-02-96

TIME

: 12:21 Hr

°C

: 17.9

pH

: 7.18

DATE 1/02

TIME 12:21

SAMPLE 1

°C

17.9

pH

7.18

PRINT