The LCD will show "Err5" to indicate that the electrode is now
ready for slope calibration.

• Immerse the pH electrode and the probe into

pH4.01 or 10.01 solution.

The LCD will show the corresponding buffer value
at the working temperature.

••

•• Wait approximately 30 seconds, and then press

the CON key to confirm the calibration.
The printer will confirm the calibration process
is complete

The calibration is now complete.

4 cm

(1½

")

HI 7004

ºC

pH

ºC

pH

°C

°C

Calibration com-

pleted

CON