beautypg.com

Hanna Instruments HI 9017 User Manual

Page 41

background image

41

5

PGM

PROGRAM # 4 (HI9318)

This program logs and prints pH/mV and temperature values at a
preset interval between 1 minute and 24 hours.

To enter the program press the 4
key and the PGM key.

The printout defines functions
A, B and C:
A is a start time,
B is the time interval,
C is the finish time.
Hours and minutes must be
separated by a decimal point.
The minimum time interval is
1 minute.
E.g. Starting time 10:20, finishing time 11:30 with a printout interval
of 5 minutes.

When the instrument is in this mode and it is desired to stop before
the end time, press the 0 key plus the PGM key.

PROGRAM #5 (HI9318)

This is the same as program 4 except that function "C" allows you to
select an interrupt value for pH/mV instead of end time.
When this interrupt value is exceeded, the program prints the hour and
minute of the endpoint.
To enter the program press the 5
key plus the PGM key.

PROGRAM 4

A enter start

time

B enter time int-

erval

C enter end time

A

C

B

1

1

0

0

1

.

.

.

2

3

0

0

0

5

4

PGM

1)

2)

3)

This manual is related to the following products: