Introducing the IQS-9100 Optical Switch

12

IQS-9100

Typical Applications

Measuring Optical Switch Port Insertion Loss

The insertion loss (IL) of any switch port can be measured using a light
source, power meter, and the IQS-9100 Optical Switch.

To measure optical switch port insertion loss:
1.
Connect the light source output port to the power meter detector port

using two test jumpers linked with a bulkhead connector.

2. Store the power reading obtained as a reference value in the power

meter.

Light Source

IQS-2100

ACTIVE

Power Meter

IQS-1100

Bulkhead connector

Light source
output port

Power meter

detector port