Product-Specific Commands—Description

:SOURce[1..n]:POWer:WAVelength

Description

This command selects a wavelength when using

a dual-wavelength source instrument.

*RST sets the selected wavelength to the LOW

value.

Syntax

:SOURce[1..n]:POWer:WAVelengthUPPe

r|LOWer

Parameter(s)

SelectedSources:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

UPPer|LOWer.

UPPer, switches to the highest wavelength.

LOWer, switches to the lowest wavelength.

Example(s)

SOUR:POW:WAV LOW

Wait 3 seconds.

SOUR:POW:STAT ON

Notes

Instrument must be in the Ready state to execute

this command. This command can cause the

instrument to enter the Stabilizing state.

See Also

SOURce[1..n]:POWer:WAVelength?

SOURce[1..n]:POWer:WAVelength:LOWer?

SOURce[1..n]:POWer:WAVelength:UPPer?