beautypg.com

Source[1..n]:am:internal:frequency – EXFO IQS-2100 Light Source for IQS-500/600 User Manual

Page 64

background image

SCPI Command Reference

58

IQS-2100

Product-Specific Commands—Description

:SOURce[1..n]:AM:INTernal:FREQuency

Description

This command selects the internal modulation

frequency. The internal modulation is 50 % duty

cycle at the selected frequency.

*RST sets the modulation frequency to 0 Hz

(CW).

Syntax

:SOURce[1..n]:AM:INTernal:FREQuency

HZ]>

Parameter(s)

ModulationFreq:
The program data syntax for

is defined as a

DATA> element followed by an optional

element. The

allowed element is

HZ.

The parameter is the new

modulation frequency: 270, 1000, 2000, or 0

(CW).

Example(s)

SOUR:POW:STAT ON

SOUR:AM:INT:FREQ 2000Hz

See Also

SOURce[1..n]:AM:INTernal:FREQuency?