beautypg.com

EXFO FTB-5500B PMD Analyzer User Manual

Page 98

background image

Troubleshooting

90

FTB-5500B

Solving Common Problems

Low signal-to-noise ratio.

Power too-low in the

selected band.

³

Use a source with a

higher output power.

³

Clean connectors

carefully.

³

Ensure connections are

made properly.

³

Ensure the source

wavelength matches the

selected band for the

Polarization Mode

Dispersion Analyzer.

Possible coupling mode

selection error.

The selected coupling mode

(Telecom or

Polarization-Maintaining)

does not match the coupling

mode the Polarization Mode

Dispersion Analyzer detects.

Calculation will be

performed using the

selected coupling mode.

Ensure that the selected

coupling mode is

appropriate.

The signal is too low to

compute a PMD value. The

acquisition operation was

canceled.

³

Power too-low in the

selected band.

³

The source being used

is modulated.

³

Ensure the input power

is sufficient.

³

Use a LED or SLED

source.

The signal is saturated.

Unable to compute PMD

value. The acquisition

operation was canceled.

Power is too high.

Reduce the source output

power.

Message

Possible cause

Solution