beautypg.com

1 specification, 2 areas of use, 3 methods of use – ENMET PGD2Manual.pdf User Manual

Page 6: 4 basic precautions

background image

ENMET Corporation

PGD3-IR

3

5

I

NTRODUCTION

This Instruction Manual provides information on the General Operation of the PGD3-IR instrument.

5.1

Specification

Size

7.25” high x 4.25” wide x 1.95” deep (185mm x 108mm x 50mm) excluding case

Weight

28 oz (1Kg)

No. of Gases

Up to four.

Gas Types

Flammable, Toxic (including Carbon Dioxide), Oxygen, (See section 10 for sensor specifications).

Internal Pump

Available option, to order (suffix P). 3.5 Volts dc.

Power Sources

Rechargeable battery pack type PGD/BATT.

Minimum operating times

4 to 12 hours: Sensor configuration dependent

Visual indications

i). 2 x 16 character LCD display with backlight.

ii). large flashing red lens section used to indicate alarm conditions.

iii). Green ‘Power’ and amber ‘Fault’ indicators in lens.

Audible indication

Electromagnetic alarm in red lens section.

Alarm levels

Flammable : single level instant alarm.

Toxic (each sensor) : 3 levels - instant, STEL, TWA(LTEL).

Oxygen : 2 levels - instant low, instant high.

Automatic calculation of STEL and TWA(LTEL) readings.

Display formats

ppm, %, LEL.

Controls

i). Front panel - on/off, right arrow, left arrow, on/off/menu.

ii). Top panel - display backlight, alarm mute.

Temperature range

i). Operating: 23°F to 104°F (-5 to + 40°C)

ii). Storage: -4°F to 122°F -20 to + 50°C)

Environmental Rating

IP66.

Humidity range

15% to 95% relative humidity, non-condensing. Operation is possible in the range 0-100% R.H.
however, sensor life will be reduced by continuous operation at extremes of humidity and
temperature.

Electromagnetic
compatibility

CE Marked, Tested to EN50081-2, EN50082-1, EN55022 (Limit B).

5.2

Areas of use

The PGD3-IR Portable Gas Detector is intended to be used for personnel protection and is specifically designed and certified to be
safe in the following areas: -

i) Confined spaces

that is, areas where normal atmospheric ventilation is not available, e.g. sewers, tunnels, ducts, storage vessels etc.

ii) Hazardous areas

classified as Class 1, Div 1, Groups A, B, C, D.

5.3

Methods of use

The PGD3-IR can be utilised for: -
Personal protection

With the instrument worn on the person as close to the breathing area as possible and switched on at all times.

5.4

Basic Precautions

The following basic precautions should be observed in order to ensure that the instrument provides the protection for which it is
intended: -
a)

Ensure that the instrument is of the correct type for the gases to be detected.

b)

Before use, check that the instrument is within its calibration period - see Section 6.6.

c)

Before use, check that the instrument is not physically damaged and that all covers are intact.

d)

Before use, establish that the instrument has been charged sufficiently.

e)

When in use, ensure that the gas sensor inlets are not obstructed - the instrument must not be used whist in a pocket, bag or
other confined container.

f)

When lowering the instrument into for example, a manhole, ensure that it is not lowered into fluids.

g)

Use the instrument in its rubber boot at all times.