1 specification, 2 areas of use, 3 methods of use – ENMET PGD2Manual.pdf User Manual
Page 6: 4 basic precautions
ENMET Corporation
PGD3-IR
3
5
I
NTRODUCTION
This Instruction Manual provides information on the General Operation of the PGD3-IR instrument.
5.1
Specification
Size
7.25” high x 4.25” wide x 1.95” deep (185mm x 108mm x 50mm) excluding case
Weight
28 oz (1Kg)
No. of Gases
Up to four.
Gas Types
Flammable, Toxic (including Carbon Dioxide), Oxygen, (See section 10 for sensor specifications).
Internal Pump
Available option, to order (suffix P). 3.5 Volts dc.
Power Sources
Rechargeable battery pack type PGD/BATT.
Minimum operating times
4 to 12 hours: Sensor configuration dependent
Visual indications
i). 2 x 16 character LCD display with backlight.
ii). large flashing red lens section used to indicate alarm conditions.
iii). Green ‘Power’ and amber ‘Fault’ indicators in lens.
Audible indication
Electromagnetic alarm in red lens section.
Alarm levels
Flammable : single level instant alarm.
Toxic (each sensor) : 3 levels - instant, STEL, TWA(LTEL).
Oxygen : 2 levels - instant low, instant high.
Automatic calculation of STEL and TWA(LTEL) readings.
Display formats
ppm, %, LEL.
Controls
i). Front panel - on/off, right arrow, left arrow, on/off/menu.
ii). Top panel - display backlight, alarm mute.
Temperature range
i). Operating: 23°F to 104°F (-5 to + 40°C)
ii). Storage: -4°F to 122°F -20 to + 50°C)
Environmental Rating
IP66.
Humidity range
15% to 95% relative humidity, non-condensing. Operation is possible in the range 0-100% R.H.
however, sensor life will be reduced by continuous operation at extremes of humidity and
temperature.
Electromagnetic
compatibility
CE Marked, Tested to EN50081-2, EN50082-1, EN55022 (Limit B).
5.2
Areas of use
The PGD3-IR Portable Gas Detector is intended to be used for personnel protection and is specifically designed and certified to be
safe in the following areas: -
i) Confined spaces
that is, areas where normal atmospheric ventilation is not available, e.g. sewers, tunnels, ducts, storage vessels etc.
ii) Hazardous areas
classified as Class 1, Div 1, Groups A, B, C, D.
5.3
Methods of use
The PGD3-IR can be utilised for: -
Personal protection
With the instrument worn on the person as close to the breathing area as possible and switched on at all times.
5.4
Basic Precautions
The following basic precautions should be observed in order to ensure that the instrument provides the protection for which it is
intended: -
a)
Ensure that the instrument is of the correct type for the gases to be detected.
b)
Before use, check that the instrument is within its calibration period - see Section 6.6.
c)
Before use, check that the instrument is not physically damaged and that all covers are intact.
d)
Before use, establish that the instrument has been charged sufficiently.
e)
When in use, ensure that the gas sensor inlets are not obstructed - the instrument must not be used whist in a pocket, bag or
other confined container.
f)
When lowering the instrument into for example, a manhole, ensure that it is not lowered into fluids.
g)
Use the instrument in its rubber boot at all times.