8 alarm levels – ENMET PGD2Manual.pdf User Manual

Page 13

background image

ENMET Corporation

PGD3-IR

10

6.8

Alarm Levels

The alarm level settings for the instrument may be viewed via the menu button. These values represent the alarm set-points which
have been programmed into the instrument at the factory.
Press the menu button until, the user is prompted to press the up arrow key to view the readings: -


Sequential operation of the up arrow key shows in turn the alarm set points for each sensor. A typical display sequence for a 4-gas
instrument would be:



or FLAM 1.00 %VOL if in the volume mode

N

OTE

: Some units are dual range LEL and % vol.

Alarm setting for the Flammable sensor - one setting




For Example

:





Alarm settings for toxic sensor, Hydrogen sulphide - three settings

For Example

:





Alarm settings for the Oxygen sensor - two settings




For Example

:





Alarm settings for infrared sensor, Carbon Dioxide - three settings


Important: - The Flammable sensor alarm setting cannot be set above 60% LEL.

ALARM LEVELS
PRESS [ ] KEY

INSTANT ALARM
FLAM 10% LEL

INSTANT ALARM
H

2

S 10 ppm

STEL ALARM
H

2

S 15 ppm

LTEL ALARM
H

2

S 5 ppm

INSTANT ALARM
O

2

LOW 19.5%

INSTANT ALARM
O

2

HIGH 23.5%

INSTANT ALARM
CO

2

2000 ppm

STEL ALARM
CO

2

5000 ppm

LTEL ALARM
CO

2

5000 ppm