beautypg.com

Static measurements, Dynamic measurements – Elenco AM Radio Kit User Manual

Page 26

background image

-25-

Figure 23

Figure 22

STATIC MEASUREMENTS

With the power turned OFF, connect the VOM to your
circuit as shown in Figure 22. Connect a clip lead from
test point two (TP2) to the collector of Q1. This short
prevents Q1 from oscillating. Set the VOM to read 2
volts DC accurately and turn the power ON. The DC

voltage at TP1 should be 1.6 volts. If the voltage in
your circuit differs by more than 0.5 volts, leave the
power ON and check the battery voltage. If the battery
voltage is greater than 8.5 volts, turn the power OFF
and check components R1, R2, R3 and Q1.

Q1 BIAS

If you do not have an oscilloscope, go to the Final Alignments With No Test Equipment Section.

DYNAMIC MEASUREMENTS

With the power turned OFF, connect the oscilloscope
to the circuit as shown in Figure 23.
Set the oscilloscope for a vertical sensitivity of 1
volt/division and turn the power ON. The oscilloscope
should display a low voltage sine wave. The

frequency of the sine wave should change when
capacitor C1 is turned. If your circuit fails this test,
turn the power OFF and check components Q1, C1,
C2, C3, L1 and L2.

OSCILLATOR CIRCUIT

If you do not have an RF generator, go to the Final Alignments with No Test Equipment Section.

V

Ohms

COM

V

Oscilloscope

TP8

TP8

Clip Lead