Calibration 51

6.2 VR

Assignments

There are five variable resistors (VR) on the ACL-8111 board to allow
you making accurate adjustment on A/D and D/A channels. The function
of each VR is specified below:

VR No.

Function

VR1

D/A full scale adjustment

VR2

D/A zero offset adjustment

VR3

A/D offset adjustment

VR4

A/D full scale adjustment

VR5

A/D programmable amplifier offset adjustment

6.3 D/A

Calibration

The D/A input reference voltage should be connected to for calibration. You
may use the JP1 to select the -5V or -10V reference. The zero offset and full
scale of D/A channel can be turned through VR1 and VR2. The full-scale
gain of D/A is adjusted by VR1, and zero offsets are adjusted by VR2.

Note: A precision voltmeter to measure the D/A output is very important.

The procedures of calibration are as follows:

Full scale adjustment :

1. Set JP1 to select -5V as reference voltage.

2. Connect VDM (+) to CN1 pin 30 and VDM (-) to A.GND.

3. Write digital data '4095' (0x0FFF) into D/A's digital data Register

(BASE+4 and BASE+ 5).

4. Adjust VR1 until your voltmeter's voltage reading is equal to +5V

(reference voltage with opposite sign).

Offset zero adjustment:

1. Write digital data ' 0' (0x0000) to D/A's digital data register (BASE+4 and

BASE+ 5).

2. Adjust

VR2 until the your voltmeter's voltage reading is 0V.