beautypg.com

Configuration – VACUUBRAND DCP 3000 User Manual

Page 35

background image

page 35 of 64

Configuration

In the ”Configuration” menu the device parameters are preselected.

Preselections
+

Use the selection knob to select the parameters.

Switch the DCP 3000 on.

Press the selection knob.

Turn the selection knob to choose a menu and confirm by pressing the

selection knob.

’’Configuration’’

Settings for

• Adjustment (of the pressure transducer)

• RS-232 (serial interface)

• Sensors (configuration and switching between several pressure trans-

ducers)

• Display (brightness and contrast of the display, language, sound,.....)

• Setpoint (valves, digital I/O interface)

• Log Data (Off, 1 . . . 3600 seconds)

• Graphic

+

Adjustment: Adjustment of the pressure transducer under vacuum

and/or at atmospheric pressure. See also sections, “Readjustment of

VSK 3000”, pg. 54. / “Readjustment of VSP 3000”, pg. 56.

Adjustment is carried out at two adjustment points: at atmospheric

pressure and under vacuum.

If no adjustment is possible in the actual pressure range, ”---- Torr” is

displayed.

+

RS-232: Configuration of the interface, setting of parameters and com-

mands, see section “Interface parameters”, pg. 43.

Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”,

”7-O-1” or ”7-E-1”, Handshake on ”no”, ”Xon-Xoff” or ”RTS-CTS” and

remote on ”On” or ”Off”.

Automatic sending of pressure values ”off” (only on demand) or in a

time interval of 1 .... 3600 seconds.