Configuration – VACUUBRAND DCP 3000 User Manual
Page 35
Configuration
In the ”Configuration” menu the device parameters are preselected.
Preselections
+
Use the selection knob to select the parameters.
➨
Switch the DCP 3000 on.
➨
Press the selection knob.
➨
Turn the selection knob to choose a menu and confirm by pressing the
selection knob.
’’Configuration’’
Settings for
• Adjustment (of the pressure transducer)
• RS-232 (serial interface)
• Sensors (configuration and switching between several pressure trans-
ducers)
• Display (brightness and contrast of the display, language, sound,.....)
• Setpoint (valves, digital I/O interface)
• Log Data (Off, 1 . . . 3600 seconds)
• Graphic
+
Adjustment: Adjustment of the pressure transducer under vacuum
and/or at atmospheric pressure. See also sections, “Readjustment of
VSK 3000”, pg. 54. / “Readjustment of VSP 3000”, pg. 56.
Adjustment is carried out at two adjustment points: at atmospheric
pressure and under vacuum.
If no adjustment is possible in the actual pressure range, ”---- Torr” is
displayed.
+
RS-232: Configuration of the interface, setting of parameters and com-
mands, see section “Interface parameters”, pg. 43.
Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”,
”7-O-1” or ”7-E-1”, Handshake on ”no”, ”Xon-Xoff” or ”RTS-CTS” and
remote on ”On” or ”Off”.
Automatic sending of pressure values ”off” (only on demand) or in a
time interval of 1 .... 3600 seconds.