Ambient conditions, Notice – VACUUBRAND DCP 3000 User Manual
Page 15
• Check the power source and the equipment’s rating plate
to be sure that the power source and the equipment
match in voltage, phase, and frequency.
• When working with residues, aggressive or condens-
able media, install a gas washing bottle if necessary.
The vacuum gauge is equipped with a short-circuit-proof
wide-range power supply with an integrated overload pro-
tection.
Assemble and lock the suitable power plug (included in
shipment) to the wall power supply prior to use.
Avoid high heat supply (e.g., due to hot process gases).
Connect the pressure transducer to the vacuum applica-
tion.
Position the pressure transducer and its vacuum line in
such a way that condensate cannot flow towards the pres-
sure transducer.
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
Comply with all
applicable and relevant safety require-
ments (regulations and guidelines). Implement the re-
quired actions and adopt suitable safety measures.
Ambient conditions
➨
Do not reach for this product if it has fallen into liquid.
There is a risk of deadly electrical shock. Unplug the
system immediately.
+
Do not use this product in an area where it can fall or
be pulled into water or other liquids.
NOTICE