Operating conditions, Notice – VACUUBRAND DCP 3000 User Manual
Page 16
• Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment
outdoors, or in an atmosphere of conductive pollution
or external condensation.
• Do not operate this product near flames.
To the best of our knowledge the equipment is in com-
pliance with the requirements of the applicable EC-direc-
tives and harmonized standards (see ”Declaration of Con-
formity”) with regard to design, type and model. Directive
EN 61010-1 gives in detail the conditions under which the
equipment can be operated safely (see also IP degree of
protection, “Technical data”, pg. 21).
Operating conditions
➨
This device is not approved for operation in potentially
explosive atmospheres.
Do not operate the device in
potentially explosive atmospheres.
➨
Devices without the
”
`
”
mark on the rating plate
are
not approved for operation with dangerous or
explosive gases or with potentially explosive or in-
flammable substances. Do not operate the device
with dangerous or explosive gases or with poten-
tially explosive or inflammable substances.
➨
Devices bearing the
”
`
”
mark on their rating plates
are
approved for operation with potentially explo-
sive atmospheres according to their classification
II 3G IIC T3 X according to ATEX, but they are not
approved for operation in potentially explosive at-
mospheres (see section “
Equipment marking (ATEX)”, pg. 19)
.
• Ensure that the materials of the vacuum gauge’s wet-
ted parts are compatible with the substances in the
vacuum system, see section “Technical data”, pg.
NOTICE