beautypg.com

Setting up and installing the equipment, Notice – VACUUBRAND DCP 3000 User Manual

Page 14

background image

page 14 of 64

+

The equipment is designed for operation at

ambient

temperatures and gas temperatures at the pressure

transducer between +50°F and +104°F (+10°C and

+40°C), or for short periods up to +176°F (+80°C) at

the pressure transducer (gas temperature). Periodical-

ly check maximum temperatures if installing the equip-

ment in a cabinet or a housing. Make sure ventilation

is adequate to maintain recommended operating tem-

perature. Install an external automatic ventilation sys-

tem if necessary. If processing hot gases, make sure

that the maximum permitted gas temperature at the

pressure transducer is not exceeded (see “Technical

data”, pg. 21).

• Ensure that the wetted parts are chemically resistant

to the processed substances prior to operation.

Use the equipment

only as intended, that is, for mea-

surement of vacuum in vessels designed for that purpose.

Any other use will automatically invalidate all warranty and

liability claims.

Remain aware of safety and risks.

Setting up and installing the equipment

+

Do not permit any

uncontrolled pressurizing (e.g.,

make sure that pipelines cannot become blocked) to

avoid a

risk of bursting!

+

Keep the VACUU•BUS cables away from heated sur-

faces.

• Comply with

maximum permissible pressures at the

pressure transducer. See section “Technical data”, pg.

21.

• Connect hoses gas tight at the pressure transducer.

• Secure hoses at hose nozzles (e.g., with hose clamp)

to prevent their accidental slipping. Ensure the stability

of hose connections.

NOTICE