Troubleshooting – VACUUBRAND Exhaust waste vapour condenser Peltronic User Manual
Page 34
page 34 of 41
Troubleshooting
Fault
Possible cause
Remedy
❑ LED at the con-
denser is off, no fan
noise.
➨ Exhaust waste vapor
condenser Peltronic™
not switched on or con-
troller switched off?
✔ Switch on condenser and/
or controller.
➨ Electrical power cord
not plugged in, electri-
cal supply failure?
✔ Plug in power cord.
Check fuse.
➨ Exhaust waste vapor
condenser Peltronic™
defective?
✔ Contact local distributor.
❑ LED is flashing
orange, no change
to continuous green
light.
➨ Very high condensation
power?
✔ Check process parame-
ters (e.g., reduce pump-
ing speed of the vacuum
pump, increase cooling
at the vacuum side of the
system).
➨ Ambient temperature
too high?
✔ Comply with maximum
permissible ambient tem-
perature.
❑ Continuous red
light at the con-
denser.
➨ Ambient temperature
much too high?
✔ Comply with maximum
permissible ambient tem-
perature.
➨ Very high condensa-
tion power and ambient
temperature too high?
✔ Check process parame-
ters (e.g., reduce pump-
ing speed of the vacuum
pump, increase cooling
at the vacuum side of
the system). Comply with
maximum permissible
ambient temperature.
➨ Exhaust waste vapor
condenser Peltronic™
defective?
✔ Contact local distributor.