Technical data, Exhaust waste vapor condenser peltronic – VACUUBRAND Exhaust waste vapour condenser Peltronic User Manual
Page 22
page 22 of 41
Technical data
We reserve the right for technical modification without prior notice!
Exhaust waste vapor condenser Peltronic™
Cooling capacity at 70°F (21°C)
ambient temperature
W
approx. 50
Control temperature
°F (°C)
50 (10)
Maximum permissible tempera-
ture of gaseous media*
°F (°C)
continuous operation: 104 (40),
for short periods (less than 5 minutes):
up to 176 (80)
Permissible ambient temperature
at storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Permissible relative atmospheric
moisture during operation
(no condensation)
%
30 to 85
Maximum permissible installation
altitude above mean sea level
ft
(m)
6500
(2000)
Permissible range of supply
voltage (including tolerance)
90-126 V~ / 180-254 V~
50-60 Hz
Rated current at:
100-120 V~ 50-60 Hz
200-230 V~ 50-60 Hz
A
A
1.6
0.7
Power draw (controlled)
W
7 - 160
Heat dissipation including
condensation heat
W
0 - 200
Permissible range of inlet and
outlet pressure (absolute)
psi
(bar)
13 to 16
(0.9 to 1.1)
Degree of protection IEC 529
IP 20
Inlet
connection for PTFE tube
I.D. 3/8”, O.D. 5/16” (10 / 8 mm)
Outlet
connection for PTFE tube
I.D. 3/8”, O.D. 5/16” (10 / 8 mm)
or hose nozzle for tubing I.D. 3/8”
(hose nozzle DN 10 mm)
Dimensions L x W x H
approx.
in
(mm)
6.9 x 7.0 x 15.4
(175 x 179 x 392)
Weight
lbs. (kg)
9.5 (4.3)
Volume of catchpot
quarts
(ml)
0.52
(500)
* if operating with potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)