spheres.
Do not operate the exhaust waste vapor
condenser Peltronic™ in potentially explosive at-
mospheres.
➨
Equipment without the
”
`
”
mark on the rating plate
is
not approved for the condensation of potentially
explosive atmospheres. Do not condense poten-
tially explosive atmospheres with this equipment.
➨
Equipment bearing the
”
`
”
mark on its rating plate
is
approved for the condensation of potentially ex-
plosive atmospheres according to its classification
II 3G IIC T3 X according to ATEX, but it is not ap-
proved for operation in potentially explosive at-
mospheres (see section
”
`
Important information:
Equipment marking (ATEX)”, pg. 20)
.
•
Do not condense substances which may form depos-
its inside the exhaust waste vapor condenser Peltron-
ic™. The exhaust waste vapor condenser Peltronic™
is not suitable for condensing substances which may
form deposits inside the condenser. Deposits and con-
densate in the condenser may lead to increased tem-
peratures even to the point of exceeding the maximum
permitted temperatures.
• Check the inlet and outlet of the exhaust waste vapor
condenser Peltronic™, if there is a danger of forming
deposits inside the condenser. Inspect the exhaust
waste vapor condenser Peltronic™ regularly and clean
if necessary. See section ”Shutdown”, pg. 32).
•
Consider interactions and chemical reactions of
the condensing media. Ensure that the materials of
the condenser’s wetted parts are compatible with the
processed substances, see section ”Technical data”,
When changing the substances processed, we rec-
ommend purging the exhaust waste vapor condenser
Peltronic™ with air or inert gas prior to changing the