Prior to use, Notice – VACUUBRAND Exhaust waste vapour condenser Peltronic User Manual
Page 29
page 29 of 41
Prior to use
+
Maximum ambient temperature: 104 °F (40 °C)
Attention: High ambient temperature reduces the cool-
ing power of the condenser. Take into consideration
heating-up of the ambient air due to the vacuum pump
and the exhaust waste vapor condenser Peltronic™.
+
Check the maximum temperatures, if installing the
condenser in a cabinet or a housing, or if the ambient
temperature is elevated. Make sure ventilation is ad-
equate.
+
Assemble exhaust waste vapor condenser Pel-
tronic™ only at the outlet of a vacuum system (at
atmospheric pressure).
• Reduce the transmission of vibration. Prevent mechan-
ical load due to rigid pipelines. Insert elastic hoses or
flexible elements as couplings.
Note: Flexible elements will compress or flatten when
evacuated if not designed for use under vacuum.
• If the equipment is installed at an altitude of more than
6500 ft (2000 m) above mean sea level, check compat-
ibility with applicable safety requirements, and adopt
suitable measures. There is a risk of overheating due
to insufficient cooling.
• If a power failure constitutes a potential source of dan-
ger, take appropriate safety measures.
Use connecting hoses with large diameter and keep them
as short as possible to avoid flow losses.
When assembling, ensure
tightness. After assembly,
check the whole system for leaks.
Keep a minimum distance of 2 in (5 cm) between fans
and surrounding items (e.g., housing, walls, etc.), or else
install an external automatic ventilation system.
NOTICE