Warning, Cleaning & disinfection, Inspections & maintenance – Oakworks Spine Positioning System II User Manual
Page 18
15
CleAnIng & DISInfeCTIOn / InSPeCTIOnS & MAInTenAnCe
CleAnIng & DISInfeCTIOn
Before cleaning with any liquid cleaner be sure to unplug the power cord from the outlet.
WARnIng
Use a 10% sodium hypochlorite (bleach) solution or Recommended Disinfectants on all surfaces. Clean all sides
of each upholstered section. Follow the directions on the disinfectant and wipe off excess.
Recommended Disinfectants
Protex, MadaCide, Accell TB, Virox®
Note: Damage caused by unapproved substances will not be covered under the warranty.
DO NOT use citrus based cleaners or other strong cleaners, such as alcohol, acetone, higher concentrations of
bleach or other products that contain high concentrations of these substances.
DO NOT expose the fabric to temperatures below 50°F/10°C or above 104°F/40°C.
DO NOT expose the fabric to direct sunlight, adhesives, liquids, or abrasive materials.
InSPeCTIOnS & MAInTenAnCe
Inspect Torso Support Pad monthly to be sure that the foam has not lost shape or firmness to the extent that patient support would
be compromised.
Inspect the base and components monthly to ensure that they have not been damaged. Replace any damaged or worn components.
Inspect face rest platform locking mechanism weekly. Use the following procedure:
Step 1 - Lock the platform cam
Step 2 - Rock platform up & down
Gently rock platform up and down and
note any “looseness” (some flexing is
normal). Look for gaps between the
aluminum parts. If you feel “loose-
ness” or see gaps, see Face Rest
Platform Cam Tightening.
Inspect joints:
Bad (gap)
Good (no gap)