Directions for use – Oakworks Marina User Manual
Page 10

The level indicator should read at least 75% of stored power for proper use. If less than this amount is
available, recharge the battery as described in the Battery Care section below.
Normal Battery Level
Battery recharge needed
battery Care:
In order to maintain the quality and ensure the life of the Marina
™
, here are some preventative steps
that can be taken to protect the battery.
• The charger must be used in a dry environment.
• Normal use of the lift function of the table is a 10% duty cycle (i.e. 2 minutes of continuous use
followed by 18 minutes off) The recommended battery recharge interval for the Marina
™
table is
every 3 hours of normal use. At this rate the battery should last for 1200 recharge cycles or 3600
usage hours.
• The battery should be recharged when the level indicator on the control box is at 75% (before it
drops to 50%). In this way, maximum battery life may be achieved. If used every day in this man-
ner, a battery should last up to two (2) years.
• It is recommended the battery be fully charged before first use after receiving the table new.
Charging time is 4 hours.
• It is recommended that the battery be left to charge over periods of non-use and that only a freshly
charged battery be used after periods of non-use. Non-use is defined as overnight or over a weekend.
• DO NOT fully discharge the battery before recharging. Completely discharging the battery before
recharge will shorten the recharge cycles to that of approximately 180 recharge cycles and thus
shorten the life of the battery.
You can receive the most from the battery on your Marina
™
by following these easy recommendations
and care instructions.
7
DIReCTIOnS fOR USe
Recharging the battery:
Pull the battery out and place it firmly onto the charger bracket. Plug in the charger to begin charging the
battery.