beautypg.com

Warning, Danger, Important safety instructions – Oakworks 121A Procedure Chair Rev.D User Manual

Page 5

background image

2

IMPORTANT SAFETY INSTRUCTIONS

The chair should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the chair should

be observed to verify normal operation in the configuration in which it will be used.
The chair is designed to be a stand alone chair. This chair must not be modified or incorporated into any other equipment.
As with any moving mechanism there are potential pinch points around and underneath the chair. It is the responsibility

of the operator of this equipment to insure that bystanders are not in the area below or around this equipment during opera-

tion.
Proper operation of this equipment is very important for the safety of the operator, patient, and any other individuals in the

area of this equipment. Directions for use of this equipment are described in this manual. The operator should read these sec-

tions carefully.
Weight Limit: (patient and accessories) 550 lbs./250 kg. Do not exceed.
Be certain that the chair is completely lowered without any tilt being present prior to discharging an ambulatory patient.

The patient may lose balance and fall.

This chair is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
Do not lift the chair by the decorative skirt. This will can damage the cahir.
When lowering the chair or using the Trendelenburg functions, make sure there is nothing underneath the table top that

can impede motion (like stools, cabinets, accessory parts, cleaners, etc.).

Use this furnishing only for its intended use as described in these instructions. Do not use attachments not recommended by

the manufacturer.
Close supervision is necessary when this furnishing is used near children or disabled persons.
The Oakworks Procedure Chair have two lifting columns that raise and lower the chair. Some chairs have one or two

actuators that allow the client to be positioned for both client and therapist comfort. Each of these devices has a duty

cycle of approximately 10%, meaning 2 minutes of operation within a 20 minute period. Exceeding this duty cycle can

damage this equipment.

WARNING

To reduce the risk of burns, fire, electric shock or injury to persons:
1. Unplug this furnishing from the electrical outlet before cleaning.
2. Unplug from outlet before adding or removing parts.
3. Never operate this furnishing if it has a damaged cord or plug, if it is not working properly, if it has been dropped or

damaged, or dropped into water. Contact Oakworks Customer Service before use.

4. Keep the cord away from heated surfaces.
5. Never drop or insert any object into any opening.
6. Do not use outdoors.
7. Do not operate where aerosol (spray) products are being used or where oxygen is being administered.

DANGER

For procedures where accidental table motion is a safety concern, the foot control must be locked out.
Risk of electric shock - Connect this furnishing to a properly grounded outlet only. See Grounding Instructions in this

manual.
Electrical Shock Hazard. The power supply/control module is located under the chair. No user serviceable parts are inside the

control box. Refer servicing to qualified personnel. Unplug wall plug prior to contact with any cables connected to the

power supply.