Т³гж 7, Specifications – LABEC AA-17BS User Manual
Page 6
Operation manual
SEA Series Steam Sterilizers
Figures
2.Specifications
Chamber inner dimensions (Diameter X Depth)..........................................22L:249mm x 450mm
17L:249mm x 355mm
Power supply ..................................................................................
According to the nameplate
Sterilizing temperature /pressure Bar(Under standard atmospheric pressure)..121
/(1.0
1.3)Bar
...........................................................................................................................134
/(2.1
2.3)Bar
Power fuse...............................................................................................................................T12A
Water tank capacity.....................................................................................................................3.5L
2
Cautions
Standard atmospheric pressure:
It is the atmospheric pressure at sea level. It refers to 1 Standard atmospheric pressure.
The value is 0.1Mpa=1000Hpa=1.0 Bar. Atmospheric pressure decreases by higher altitude.
For instance: at or lower than 3000 meters , An increase of altitude of 1000 meters will result in a decrease of
ATM PR .by 0.1Bar approximately.
Note: Temperature/Pressure here means the chamber temperature and pressure during the phase
of sterilization. These value vary under different atmospheric pressure.
Tip:
Operation temperature..............................................................................................+5
+40
Relative humidity range...............................................................................................30% 75%
The maximum wight of instruments which load into the sterilizer chamber during one cycle.
Unwrapped cycle 121
22L/17L: 6.5kg
Unwrapped cycle 134
22L/17L: 6.5kg
wrapped cycle 121
22L/17L: 3.2/2.5kg
wrapped cycle 134
22L/17L: 3.2/2.5kg
Package select
Start/stop
Water outlet
Refer to the operation
manual
Test cycle select
Last cycle select
B type device
Note of high
temperature
Program select
Fast sterilizing select
Grounded correctly
Distilled water only