Т³гж 23 – LABEC AA-17BS User Manual
Page 22

Operation manual
SEA Series Steam Sterilizers
18
Over time small impurities will deposit on the drain filter, and this moy affect vacuum and
drying. In order to prolong the life of the drain filter please use the following cleaning
procedure (see fig 9-5):
-5
Fig9
9.4 Change fuse (Fig 9-6)
-6
Fig9
Please check the O ring of Drainning Filter during cleaning. If this part is broken please replace
it with O ring in the accessories supplied.
9.3 Clean the drain filter
Unplug the autoclave before changing the fuse
Caution
1.Unscrew the fuse holder with a screw driver counter clockwise.
2.Make sure to replace the faulty fuse with a correct fuse.
3.Put back the fuse holder and screw it on clockwise with a screw driver.
1.Locate the water filter.
2.Rotate off the filter screw
3.Pull off the rubber filter from the connection
4.Take off the filter from the blue rubber.
5.Clean the filter or replace with a new one.
6.After you clean or change a new filter shim,
plug in the water filter back on the chamber
connection.
Note: please turn the blue rubber to face down
hard to touch the chamber and tighten the screw
nut.
Note