SEA Series Steam Sterilizers
Operation manual
23
13.Appendix.
Form 1 Work mode chart
Form 2 Error code chart(In case of faults, please contact your distributor or the immediately)
Sterilizing temperature
Sterilizing pressure
Class
Pre-heating
Vaeuuning time
Sterilizing time
Drying time
Quick drying time
Total time
B cycle
134
B&Dtest
Admission time
134
134
121
134
2.10
2.10
2.10
1.10
2.10
B
B
S
B
B
0-7min
0-7min
0-7min
0-7min
0-7min
10min
10min
10min
10min
4min
4min
5min
5min
5min
5min
5min
4min
18min
4min
16min
3.5min Keep pressure
10min
10min
10min
10min
10min
1min
5min
5min
5min
5min
36(31)min 50(45)min
30(25)min
48(43)min
19.5min
15min
1
2
3
4
5
6
7
8
Item
9
Error
sensor error
super pressure >2.6Bar
o
super temperature >145 c
Constant temperature fails
block of steam pipe (20S<0.3Mba)
failure of the door
Steam generator over
temperature (>230 C)
heating overtime (>25 min)
pressurization
overtime (>60 min)
Solution
Examine inner temperature
sensor or pressure sensor
examine outer temperature
sensor
Return parameter autoadjus
programm
Examine the steam output
valve or vacuum valve
examine the door switch
Check steam generator
temperature sensor
examine the electronic connection
of heat ring and heat stick
examine the steam input pipe
examine sensor
18min
Bcycle
Unwrappd
cycle
RUB & PLAS
cycle
Vacuum test
Clean test
Clean 20
min
10
min
30
min
Drying
The figures in the chart are only for reference, the sterilization effect verified by user, and the user can
choose the sterilization time according to the verification results. Which cycle should be adopted is
according to the type of the instruments.
Note: